Labtone

Medium-force shaker

2000 kgf - 3200 kgf

The EV series of electromagnetic vibration test systems simulate the vibration environment under laboratory conditions, testing various vibration test applications’ shock resistance and reliability. In the laboratory, with the aid of a shaker, sinusoidal, random, and classic shock signals, as well as road patterns, can be performed. The aforementioned laboratory activity is essential for the validation of product quality.


DESCRIPTION

Labtone EV series’ shakers are specifically designed to meet the need for long-time operation. Vertical and horizontal vibrations can be achieved by installing a vibration-resistant basis. The standard platform is equipped with high-efficiency dampers to minimize vibrations to the building. An additional basis is not needed in most cases.


FEATURES

  • Robust suspension system and linear motion guiding, strong carrying capacity, high stability
  • Load center damper has high static stiffness and low dynamic stiffness, strong carrying capacity, perfect performance on amplitude variation
  • High-efficiency power switching provide the finest power consumption and minimize harmonic distortion
  • Quick self-diagnosis with safety interlock, high safety reliability
  • Airbag shock isolation device for vibration platform without the need for additional foundation, perfect reproduction of vibrational wave, and reduce vibration transmittance
  • Provide horizontal and vertical expansion platforms for different application

Model EV220 EV320 EV222 EV222B EV322 EV422 EV232 EV232B EV432
Amplifier model LA20K LA30K
Shaker model A20S32 A20T32 A20S34 A20S44 A20T34 A20F34 A30S34 A30S44 A30F44
Rated sine force (KN/Kgf) 20/2000 24/2400 22/2200 32/3200
Rated shock force (KN/Kgf) 40/4000 45/4500 44/4400 64/6400
Frequency range (Hz) 5-2700 5-2500 5-3000 5-2800 5-3000
Max acceleration sine/random (g) 98/60 85/60 100/60 90/60 70/60 100/60 85/60
Max velocity (m/s) 2.0
Max displacement P-P (mm) 51 76 51 76 100 51 100
Max payload (Kg) 300 400 500
Armature diameter (mm) 320 342 442 342 442
Resonance frequency (Hz)(± 5%) 2200 2600 2700 2400 2500
Effective moving mass (Kg) 24 25 30 35
Blower B752
Blower power (KW) 7.5
Blower volume flow rate/Pressure (m3/min; KPa) 41;2.6
System power (KVA) 42 56